Argentum Solutions, Inc.

    Sterling guidance on corrosion and materials degradation


 

Potential-pH Diagrams
THERMEXPERT - Potential-pH diagram generator

Intelligent Tools

POLEXPERT - Polarization Scan Artificial Neural Network Expert System

SEQEXPERT - Sequential Immersion Test Artificial Neural Network Expert System

CYLEXPERT - Rotating Cylinder Electrode Intelligent Rotation Rate Calculator

Corrosion Calculator

Corrosion Rate Calculator


Corrosion Economics Estimator

FINCALCULATOR - Corrosion Economic Calculator


TUTORIAL ON ELECTROCHEMICAL IMPEDANCE SPECTROSCOPY

David C. Silverman


Table of Contents

Overview of Tutorial
Overview of EIS
Analysis Using Simple Circuits
Constant Phase Element
Corrosion Rate Estimation
        Two Capacitive Relaxation Time Constants
Critical Criteria for Proper Spectra
Diffusion Impedance
Pseudoinductance

Diffusion Impedance

One type of impedance spectrum that is sometimes observed when examining such phenomena as corrosion in soils, concrete, or stagnant water is shown in this figure . The supposition is that the linear portion of the response continues on to infinity in real and imaginary coordinates as the frequency approaches zero. This behavior is often misnamed the Warburg impedance. This structure is actually a subset of diffusion influenced impedance. It should be called the infinite length Warburg impedance to differentiate it from the more general finite-length Warburg solution. The infinite length Warburg impedance is derived from a solution of the diffusion equation in which migration is into a semi-infinite medium. This situation is somewhat rare because in many practical systems diffusion occurs over finite distances. In addition, note that one of the criteria for the frequency spectrum to be called an impedance is that the spectrum must be finite valued at infinity. One way for this criterion to be fulfilled is for the equivalent circuit to be an open circuit or a capacitor at infinite distance. One should exhaust other possibilities before stating that the situation is an infinite length Warburg impedance.

If one assumes that the Nernst diffusion-layer thickness is comparable to the distance traveled by diffusing species while subjected to low frequency oscillating perturbations then the following equation can be derived for the Warburg diffusion contribution to the impedance:

                                                              (20)

This solution is often appropriate for a stirred electrolyte or a rotating electrode, environments often encountered in practical corrosion testing. Incorporating this finite length Warburg impedance along with a relaxation "RC" time constant results in the following analogous circuit . This circuit was simulated by using a time constant of 1 second, a solution resistance of 10 ohm-cm2, a charge transfer resistance of 104 ohm-cm2 and a CPE exponent of 0.8. The result is shown in this figure . The solid line corresponds to the portion of the spectrum terminated at a low measurable frequency of between 10-4 and 10-3 radian/s. This portion of the curve has the appearance of an infinite Warburg impedance . In practice, before concluding that a spectrum corresponds to an infinite length Warburg impedance, the corrosion practitioner should ensure that the structure is not caused by terminating the spectrum at too high of a frequency. Addressing that question may be difficult because sometimes the measurement cannot proceed to low enough frequencies to extract the entire spectrum. This point can be seen by transforming the above simulation to Bode format . As shown in this figure, most experimenters do not or cannot obtain spectra below about 10-4 to 10-3 radian/s. For example, a frequency of 10-7 radian/s would take the better part of a year to complete. The key point is that before one assumes an infinite Warburg impedance, one must fully examine all other possible diffusion analyses. Further information can be found in J. R. Macdonald, Impedance Spectroscopy, John Wiley & Sons, NY, 1987.




Next Page: Pseudoinductance

Previous Page: Critical Criteria for Proper Spectra

Return to Table of Contents







David C. Silverman, Ph.D. - Primary Consultant
E-Mail:     dcsilverman@argentumsolutions.com
Phone:     314-576-3586
Fax:         314-754-9825
Address:   The Argentum House
                14314 Strawbridge Ct.
                Chesterfield, MO 63017